Specifications
Number of Channels
1
Compatible Detectors
Supports photodiode, thermal, irradiance and dosage sensors
Communication Interfaces
RS232
Data Transfer
500Hz @ baud rate 115200
Analog Output
User selected 1v, 2v, 5v, or 10v full-scale; 0.03% resolution. 100 ohms output impedance Two modes: Digital and Raw
Accuracy
±0.25% ±20pA for thermopiles and photodiodes ±0.25% for pyroelectrics
Analog Input Range
15 nA to 5.0 mA full scale in 16 ranges
Dynamic Range
9 decades
Sampling Rate
Photodiode and Thermopile: 15Hz Pyroelectric: 500Hz
Photodiode Measurement
Average power, Exposure
Thermopile Measurement
Average power, single shot energy, pulsed power
Pyroelectric Measurement
Pulse energy (every pulse up to 500Hz), average power, exposure
Resolution
16 bits plus sign for thermopiles and photodiodes 12 bits no sign (0.025% resolution) for pyroelectrics
Noise
500nV or 1.5pA + 0.0015% of input range @3Hz for thermopiles and photodiodes 2mV for pyroelectrics
Dimensions
114 x 80 x 29 mm (LxWxH)
Power Requirements
Industrial +12V wall cube power supply included
Operating Temperature
15°C to 35°C, <70% RH, non-condensing
Storage Temperature
0°C to 50°C, <90% RH
CE Certified
Compliant
RoHS
Compliant
Weight
0.176 kg
Number of Channels
1
Compatible Detectors
Supports photodiode, thermal, irradiance and dosage sensors
Communication Interfaces
RS232
Data Transfer
500Hz @ baud rate 115200
Analog Output
User selected 1v, 2v, 5v, or 10v full-scale; 0.03% resolution. 100 ohms output impedance Two modes: Digital and Raw
Accuracy
±0.25% ±20pA for thermopiles and photodiodes ±0.25% for pyroelectrics
Analog Input Range
15 nA to 5.0 mA full scale in 16 ranges
Dynamic Range
9 decades
Sampling Rate
Photodiode and Thermopile: 15Hz Pyroelectric: 500Hz
Photodiode Measurement
Average power, Exposure
Thermopile Measurement
Average power, single shot energy, pulsed power
Pyroelectric Measurement
Pulse energy (every pulse up to 500Hz), average power, exposure
Resolution
16 bits plus sign for thermopiles and photodiodes 12 bits no sign (0.025% resolution) for pyroelectrics
Noise
500nV or 1.5pA + 0.0015% of input range @3Hz for thermopiles and photodiodes 2mV for pyroelectrics
Dimensions
114 x 80 x 29 mm (LxWxH)
Power Requirements
Industrial +12V wall cube power supply included
Operating Temperature
15°C to 35°C, <70% RH, non-condensing
Storage Temperature
0°C to 50°C, <90% RH
CE Certified
Compliant
RoHS
Compliant
Weight
0.176 kg